35-MHz 64-Element 2-2 Composite Linear Array for Ophthalmic Imaging

This 64-element linear array operates at 35 MHz for high-resolution ophthalmic imaging. The array uses a 2-2 piezo-composite fabricated by mechanical dicing of bulk Navy Type VI PZT with 14 µm kerfs filled with Al₂O₃-loaded epoxy. Elements are spaced at 50 µm pitch (1.2λ) with 3 mm elevation aperture and cylindrical epoxy lens focused at 10 mm. A 75 Ω coaxial cable provides impedance matching to 50 Ω electronics. Measured bandwidth averages 55%, insertion loss is 22.8 dB, and crosstalk is < −24 dB. Synthetic aperture imaging of a wire phantom demonstrates 42 µm axial and 95 µm lateral resolution.

Parameter

Value

Array Type

Linear array, 2–2 composite

Number of Elements

64

Center Frequency

35.3 MHz

Bandwidth (‑6 dB)

55%

Piezoelectric Material

Navy Type VI fine‑grain PZT

Composite Thickness

50 μm

Element Pitch

50 μm (1.2λ)

Elevation Aperture

3 mm

Elevation Focus

10 mm

Insertion Loss

22.8 dB

Crosstalk (adjacent)

< −24 dB

Axial Resolution

42 μm

Lateral Resolution

95 μm